User adjustments
OPTIONS OF THE MENUS
Different menus are accessible on the MAIN MENU screen.
SPE CYCLE menu
Use this menu to carry out specific procedures necessary to ensure the proper operation
of specific measurement cycles (for example, adjustment of pressure regulator).
none
Regulator adjust.
Inf Fill
Piezo auto zero
AUTO VOL
SPECIAL CYCLE MENU
Label Special cycle Description of the cycle
none
None No special cycle selected
Regulator adjust.
Regulator adjustment Adjustment of regulator in front panel
Inf Fill
Infinite fill Pressurize the part with a infinite fill time
Piezo auto zero
Piezo auto zero Auto zero cycle on the piezo sensor
AUTO VOL
Automatic volume Volume calculation for automatic program
selection
Some parameters are displayed when specific functions are activated.
Label Special cycle Description of the cycle
Custom Unit Learn
Custom unit learn Define unit on a master leak
Custom Unit Check
Custom unit check Check correct unit learning on a master leak
TO START SPECIAL CYCLES...
1. On the SPECIAL CYCLE MENU screen, select a
cycle, and press to validate.
2. Press Start to start the cycle.
3. To stop the current cycle press Reset .