QSG_F600.415.00_EN_01 / 2019-04-16
20 / 38
PNEUMATIC CONNECTORS
Pneumatic connectors used to connect the part under test are located on the back panel
of the device.
Pneumatic supply
The pneumatic supply has to meet specific requirements recommended by ATEQ. Refer to
Good practices and safety instructions section.
A specific filter may be necessary.
The air is supplied via the filter located on the back panel of the device.
Metal air filter
The metal filter is used for 1MPa
(145PSI) range.
The maximum pressure admissible is
1.2MPa (174PSI).
Plastic air filter
The plastic filter is used for 0.5MPa
(72.5PSI) range (direct and indirect
modes) or 2MPa (290PSI) range (for
pilot valves input).
The maximum pressure admissible is
690kPa (100PSI).