This message may be displayed during the coupler calibration for the RECD transducer. It indicates that the
calibration is significantly different from normal which may cause significant error while fitting.
• Check that the RECD transducer is connected.
• Check that the RECD transducer is inserted into the HA-2 coupler and the HA-2 coupler is tightly attached
to the coupler microphone.
• Check that the coupler microphone is connected.
Invalid RECD real ear measurement
Invalid RECD real ear measurement.
This message may be displayed during the RECD real ear measurement. It indicates that the measurement is
significantly different from normal which may cause significant error while fitting.
• Check that the RECD transducer is connected.
• Check that the probe module is connected. Ensure the probe tube is not pinched or blocked.
• Check that the RECD transducer is inserted to inserted into the ear that is under test.
Barcode scanner malfunction
If there is no red scanning beam when the trigger is pulled, tug on the cable where it enters the scanner. If it
comes loose easily, push it back in until it clicks in place.
If the scanning beam stays on when the trigger is released or if the barcode scanner fails to read known good
barcodes, try the following:
1. Unplug the scanner from the equipment, then reconnect it. If the problem persists…
2. Print this page on an external printer (see External display, printer, speaker) or from the electronic user’s
guide (see Electronic user’s guide).
3. If your barcode scanner has a PSC logo, scan these two barcodes, starting with the top one. This will reset
the scanner to its default settings.
129