EasyManua.ls Logo

audioscan Verifit VF-2 - Test Box Directional Procedure

Default Icon
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verifit
®
User's Guide Version 3.12
November 2015
Test box directional procedure
To test directional function in the test chamber:
1. Mount the hearing aid to the appropriate coupler (see Coupling the hearing instrument).
2. Position the hearing instrument in the test chamber as shown in 9.6: Directional ITE positioning or 9.7:
Directional BTE positioning.
3. Press <Tests>, then highlight and <PICK> [Test box measures], then highlight and <PICK> [Directional].
4. Highlight and <PICK> Test 1, 2, 3 or 4. This presents test signals from both speakers. Two real-time curves
appear on the screen - the heavier one is the response to the left speaker signal and the lighter one the
response to the right speaker signal. For a non-directional instrument, the two curves will be nearly identical.
Setting the instrument for directional operation should produce a separation of the curves.
5. On the Test control poster, highlight and <PICK> the Level window, then highlight and <PICK> a level
from the drop-down list. For an adaptive directional instrument, a stimulus level between 60 and 70 dB
should cause the separation to change over a 15 45 second period. For a noise-reducing instrument, both
curves may change with time and stimulus SPL.
6. Highlight and <PICK> [Speech S/N] within the Test control poster and select a signal to noise ratio from the
list box (defaults to 0 dB). A S/N ratio between 0 and 6 dB will usually induce directional performance after
15 45 seconds. Optionally, highlight and <PICK> [Speech S/N] and select Off. This presents noise
alternating between the left and right test speakers. Speech-adaptive directional instruments may appear to be
non-directional if [Speech S/N] is set to Off.
58

Table of Contents

Related product manuals