EasyManua.ls Logo

audioscan Verifit VF-2 - 23 Technical Specifications

Default Icon
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verifit
®
User's Guide Version 3.12
November 2015
23 Technical
Specifications
Specifications subject to change without notice
Environmental conditions
STORAGE & TRANSPORTATION
Temperature - 20°C to
+60°C
Relative humidity (non-condensing)...........5% to 95%
OPERATING
Temperature......................................+15°C to +30°C
Relative Humidity........................................20% - 50%
Atmospheric Pressure.............................81 - 106 kPa
Altitude...................................................below 2000 m
General
Power source:.........100 – 240 V, 50 – 60 Hz, 250 VA
Fuse:.......................................................T2AL , 250 V
Overall dimensions
..............................................................Rear main unit
...........................14.5x16x6.5 in. (36.8x40.6x16.5 cm)
................................................................Test chamber
..............................14x14.4x4 in. (35.6x36.6x10.2 cm)
Weight.....................................................17.6 lb (8 kg)
Display type.................fluorescent backlit active color
Display size............................................12.1” diagonal
Resolution.........................................800x600 (SVGA)
Printer................200 dots per inch thermal line printer
Paper width..................................................3” (80mm)
Headphone monitor amplifier.......1 watt into 16 ohms
Power amplifiers...............................2 @ 5 watts each
Stimulus Channels......................................................2
Measurement channels..............................................2
Connectors.......................1 - external monitor (HDMI)
...........................................................................4-USB
.......................................................1 - Ethernet (RJ45)
Test Box
Working Space.............11x5x1.5 in (28x12.7x3.8 cm)
Speaker..................2 - 2x3 in (5x7.5 cm) independent
Test box isolation @ 1kHz................................>25 dB
Induction Coils per ANSI S3.22-2003
.........1 - Telephone Magnetic Field Simulator (TMFS)
........................................1 - 23 cm diameter test loop
Battery Simulator.......................per ANSI S3.22 2003
Frequency Range..................................200 - 8000 Hz
Coupler microphone noise floor.(200 – 8000 Hz) <40 dB
SPL
Test Stimuli.....................tone, tone burst, pink noise,
..............dual-direction noise,calibrated or live speech
Test stimulus levels.....40 to 90 dB SPL in 5 dB steps
Test stimulus levels (inductive).31.6mA/m(ANSI S3.22)
Test stimulus distortion....................<2% at 90dB SPL
.....................................................<0.5% at 70 dB SPL
Test stimulus accuracy at reference mic. for tones
(200 - 2000Hz).........................................±1.5 dB SPL
(2000-8000 Hz)........................................±2.5 dB SPL
Equalization method..........real-time pressure method
Analysis frequencies per octave..............................12
Analysis filter bandwidth............................1/12 octave
Measurement accuracy at 1 kHz.........................±1db
Measurement accuracy re 1 kHz.±1 dB (200-5000 Hz)
...............................................±2.5 dB (5000-8000 Hz)
Measurement range..........................30 - 140 dB SPL
Harmonic distortion measurement..................2nd, 3rd
..............................................................or 2nd plus 3rd
Harmonic distortion range...................200 to 4000 Hz
Harmonic distortion accuracy...............................±1%
Battery drain range........................................0 - 20mA
Battery drain accuracy..........................................±5%
Battery drain resolution.................................±0.01 mA
ANSI S3.22 - 1996 and 2003 tests available
........................................................................OSPL90
..................................................................Full-on Gain
....................................................Reference Test Gain
...................................................Frequency Response
.........................................................Frequency Range
.......................................................Maximum OSPL90
......................................................Harmonic Distortion
..................................................Attack & Release time
..................................................Equivalent Input Noise
.....................................................Input/Output Curves
.............................Coupler SPL - Telephone Simulator
...................................... Simulated Telecoil Sensitivity
................................................................ Battery Drain
Other tests available
............Speechmap® real-speech audibility measures
....................Real-time adaptive directional verification
............................................Coupler SPL vs frequency
............................................Coupler gain vs frequency
..........................................................Spectral analysis
..........................................Noise reduction verification
...............................Feedback suppression verification
.................................................Distortion vs frequency
...Manual measurement of output, gain and distortion
VIEWPORT® audibility, directional, noise reduction and
feedback suppression verification
On-Ear
Speakers...................................2 - 2x 3 in. (5x7.5 cm)
Probe modules..........2 - each containing probe & ref.
microphones
Probe microphone tube.Silicone 1 mm diameter x 75 mm
Probe microphone noise floor (200 – 8000 Hz).< 45 dB
132