EasyManua.ls Logo

audioscan Verifit VF-2 - Calibrating Test Box Reference Microphone

Default Icon
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verifit
®
User's Guide Version 3.12
November 2015
Calibrating test box reference microphone
Remove any coupler from the coupler microphone before calibration.
1. Position the test box reference microphone opening concentric with the coupler microphone opening and
about 1 2 mm from as shown below.
2. Close and seal the test chamber.
3. Press <Tests>, then highlight and <PICK> [Test box measures], then highlight and <PICK> [Calibration].
4. Highlight, then <PICK> [Daily] or [Weekly] to select the calibration interval.
5. Highlight and <PICK> [Calibrate]. See Test box calibration facts
It is recommended that you print a copy of the initial calibration curve and compare it with your daily or weekly
curve. Any change in this curve indicates a drift in either the coupler or the reference microphone which needs
further investigation. See Calibration check for coupler microphone for an additional check.
42

Table of Contents

Related product manuals