Verifit
®
User's Guide Version 3.12
November 2015
The probe module should be replaced if it fails this test.
Max TM SPL setup
The Max TM SPL safety feature allows you to set a limit on the SPL, measured by the probe microphone, above
which a test terminates. The MAX TM SPL limit is shown on all on-ear measurement screens. The default limit
is 120 dB.
To choose a different Max TM SPL setting:
1. Press <Setup>, then highlight and <PICK> [Max TM SPL].
2. Highlight and <PICK> the current Max TM SPL setting.
3. Highlight and <PICK> the desired Max TM SPL setting from the drop-down list.
4. Press <Continue> to exit Setup.
This Max TM SPL setting will be in effect until you change it.