EasyManua.ls Logo

audioscan Verifit VF-2 - Page 133

Default Icon
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verifit
®
User's Guide Version 3.12
November 2015
SPL
Frequency Range................................200 to 8000 Hz
Test Stimuli.....frequency-modulated tone, tone burst,
.pink noise, dual-direction noise, calibrated or live speech
Freq. modulation..............sawtooth ±3% over 128 ms
Stimulus levels for tones.40 - 85 dB SPL in 5 dB steps
Test stimulus accuracy at reference mic. for tones
(2000- 8000 Hz)..........................................±3 dB SPL
Equalization Method.real-time modified pressure method
...............................................(stored for open fittings)
Frequencies per octave (swept tones).....................12
Frequencies per octave (tone burst)..........................3
Analysis bandwidth (speech, noise)............1/3 octave
Measurement accuracy at 1 kHz........................±1 dB
Measurement accuracy re 1 kHz.±1.5 dB (200-5000 Hz)
................................................±2.5 dB (5000-8000Hz)
Measurement Range...30-135 dB SPL (200-2500 Hz)
....................................30-140 dB SPL (2500-8000Hz)
ANSI S3.46 - 1997 tests available
....................................... Real-Ear Unaided Response
........................................... Real-Ear Aided Response
..................................... Real-Ear Occluded Response
............................................... Real-Ear Insertion Gain
Other tests available
............Speechma real-speech audibility measures
....................Real-time adaptive directional verification
...........................................On-ear harmonic distortion
...............................................On-ear spectral analysis
..............................On-ear noise reduction verification
....................On-ear feedback suppression verification
...Manual measurement of output, gain and distortion
VIEWPORT® audibility, directional, noise reduction and
feedback suppression verification
Fitting methods available
Speechmap with DSL5.0a, NAL-NL1, NAL-NL2, CAMFIT
Insertion gain with NAL-RP, NAL-NL1, Fig6, PogoII,
Berger, Libby
Sensory Loss Simulator
Simulation types...............................Linear conductive
........................Non-linear outer hair cell cochlear loss
Simulation bands......................................................65
133

Table of Contents