Test Procedure
© b2 electronic GmbH DHV0086 Rev06 71
Step Procedure (Running automatic mode)
AR7: HV Activation
Once the activation screen appears, press
the HV switch
40
within 10 seconds.
If the HV switch is not activated within 10
seconds, the “Manual Mode” screen will
reappear.
AR8: Test Start up
“Startup” appears on the screen to indicate
that the HVA is initializing test.
AR9: Test
Test begins automatically.
The timer value indicates the remaining
testing time.
The bottom line of the screen display the
preset values
AR10�1: Test End – New Phase
Display indicates end of automatic test.
For testing the next phase, select the
“Next Phase” button and push in/click the
navigation knob
41
.