MODEL 300 COMBUSTION ANALYZER
WARNINGS!
• Burn hazard. DO NOT touch the Model 300 Combustion Analyzer probe
after it is removed from a stack. Allow the probe to cool before han-
dling.
• Burn hazard. When emptying filter/trap bowl and particulate trap, be
aware that both traps may contain hot and mildly corrosive water. Take
the necessary precautions to avoid being burned.
• Burn hazard. All of the sensors contain corrosive chemicals. DO NOT
puncture or take them apart.
CAUTIONS:
• To avoid damaging the probe assembly due to excessive heat, remove
probe from stack at or before the prescribed time listed below. Allow
probe to cool to room temperature before reinserting it back into the
stack.
100–1000°F (38–538°C) ....... Unlimited
1000–1200°F (538–649°C) ....... 30 mins. per exposure
1200–1400°F (649–760°C) ....... 10 mins. per exposure
Note that the high-temperature extended-probe options have unlimited
exposure times up to 2000°F (1093°C)
• Allow probe to cool before placing it in case.
• DO NOT store the Model 300 Combustion Analyzer at temperatures
below – 4°F (– 20°C).
Instruction 24-9223 i