MODEL 300 COMBUSTION ANALYZER
Instruction 24-9223 5-15
Table 5-1. Typical NO
X
Stabilization Times
Bias removed* Stabilization Time
Less than 15 min. Less than 1 min.
Less than 1 hr. Less than 5 min.
Less than 2 days Less than 4 hr.
Greater than 2 days Up to 2 days
* Due to dead or missing batteries.
4. Reinstall the control panel and the auxiliary-display panel.
5.8 OPTIONAL DIP SWITCH SETTINGS
The settings of the analyzer’s internal switches select which sensors are
active when installed; which language is used on the printout; and which
RS-232 port settings are in effect when using an external serial device.
NOTE: Any new switch settings do not take affect until the
instrument goes through a power off/on cycle.
To set the switches, proceed as follows:
1. Remove control panel per Section 4.1 in this manual.
2. Tilt control panel upwards until the DIP switch becomes accessible
(Fig. 5-8).
Figure 5-7. Optional Bias Battery Location