Manual Configuration
0024-9486 Rev 2 27
Setup Menu Function
NOTE: The presence of AM or PM after
the time on the Set Clock display
indicates 12-hour
time format and
MM/DD/YY date format
. (This also
indicates that the instrument must be in
the North American configuration.)
Similarly, the absence of AM or PM
indicates 24-hour time format and the
date is in DD/MM/YY format (either by
default if Siegert configuration, or by
choice through the Date Format
parameter if North American
configuration).
The measured value of CO can be referenced to a
specific O
2
percentage (0% to 15%) as referenced in the
equation below.
𝐢𝑂(𝑛) =
20.9 βˆ’ 𝑂
2
π‘…π‘’π‘“π‘’π‘Ÿπ‘’π‘›π‘π‘’
20.9 βˆ’ 𝑂
2
π‘€π‘’π‘Žπ‘ π‘’π‘Ÿπ‘’π‘‘
Γ— 𝐢𝑂
Use the UP (

) and
DOWN (
ο‚‚
)
arrow buttons to
enter the O
2
reference value (n) from 0% to 15%.
Press ENTER to save the selection or ESC to revert to
the previous setting.
NOTE: The O
2
reference has a default
value of 0%. CO with respect to a 0% O
2
reference is also known as CO Air Free or
CO(0).