Manual Calibration and Maintenance
0024-9486 Rev 2 49
5.2.1. Equipment Required
Alcohol
Aerosol Can of Automotive Carburetor Cleaner
Clean Rag
Source of Compressed Air (optional)
CAUTION: Do not use flammable or combustible substances (like
carburetor fluid used for cleaning the probe) near an open flame.
5.2.2. Procedure
Step Cleaning the Probe
1 Remove gas sample hose from the top of the water trap.
CAUTION: Carburetor cleaner damages plastic components.
Take precautions not to spray cleaner onto the probe handle or
analyzer.
2 Insert the plastic spray tube of the carburetor cleaner into the gas sample
hose, and then liberally spray carburetor cleaner through the hose and out
the probe tube.
3 After spraying, remove all the residual cleaner by repeatedly flushing the gas
hose and probe tube with alcohol.
4 Wipe off the surfaces of the probe and tubing with a clean cloth.
5 Allow the parts to dry completely. If available, blow compressed air through
the probe to expedite the drying process.
6 Reconnect gas sample hose to top of the water trap.

Table of Contents

Other manuals for Bacharach Fyrite INTECH

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bacharach Fyrite INTECH and is the answer not in the manual?

Bacharach Fyrite INTECH Specifications

General IconGeneral
BrandBacharach
ModelFyrite INTECH
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals