Pulsation
level
Work step
(sec)
Idle step (sec)
1 0.1 0.9
2 0.2 0.8
3 0.3 0.7
4 0.4 0.6
5 0.5 0.5
6 0.6 0.4
7 0.7 0.3
8 0.8 0.2
9 0.9 0.1
-10
4 Use
4.1 Notes on use
Do not use any bent probes (⇒ unstable operation, loss of power).
Do not touch the sonication vessels with the oscillating probe - probes and vessels could be
damaged.
The recommended immersion depth for probes is 10 - 20 mm, to prevent the aspiration and
mixing-in of air. If mixed-in air is desired, the probe may be immersed just a few millimetres.
In the case of greater immersion depths and/or the sonication of liquids with high viscosities,
the probe is more heavily damped. As a result, it is possible that the pre-set amplitude (power)
may not be reached, especially with higher default settings (>50 %). The reason is that the HF
generator can no longer provide the required power.
The immersion depth may not exceed 20 mm in the case of micro and tapered tips.
When producing emulsions, the probe should be immersed to the level of the interface between
the liquids to be mixed.
Do not use combustible solvents in open reaction vessels since the operating safety of the
homogeniser could be compromised.
In order to shut down the device, disconnect it from the mains ( pull the plug).
17 / 4451436a GB/2017-10