Pulsation
level
Work step
(sec)
Idle step (sec)
1 0.1 0.9
2 0.2 0.8
3 0.3 0.7
4 0.4 0.6
5 0.5 0.5
6 0.6 0.4
7 0.7 0.3
8 0.8 0.2
9 0.9 0.1
-10
4 Use
4.1 Notes on use
• Do not use any bent probes (⇒ unstable operation, loss of power).
• Do not touch the sonication vessels with the oscillating probe - probes and vessels could be
damaged.
• The recommended immersion depth for probes is 10 - 20 mm, to prevent the aspiration and
mixing-in of air. If mixed-in air is desired, the probe may be immersed just a few millimetres.
In the case of greater immersion depths and/or the sonication of liquids with high viscosities,
the probe is more heavily damped. As a result, it is possible that the pre-set amplitude (power)
may not be reached, especially with higher default settings (>50 %). The reason is that the HF
generator can no longer provide the required power.
• The immersion depth may not exceed 20 mm in the case of micro and tapered tips.
• When producing emulsions, the probe should be immersed to the level of the interface between
the liquids to be mixed.
• Do not use combustible solvents in open reaction vessels since the operating safety of the
homogeniser could be compromised.
• In order to shut down the device, disconnect it from the mains ( pull the plug).