6.3 Error analysis
Errors may arise
on the plug connections
on the ultrasonic transducer
on the probes
on the HF generator
The device is robustly constructed and designed for a high level of reliability. Nevertheless,
the possibility of a malfunction due to a defective component can never be fully discounted.
Mechanical defects of the HF connector sockets, the plug connectors, the ultrasonic transducer,
etc., are possible as a result of frequent use or even incorrect handling, e.g. by dropping them.
Error analysis with HD 2070-U or HD 2200-U
Error Possible cause Remedial action
Device cannot be switched on?
Green control light does not light
up?
Is the mains voltage selector switch
on the generator set incorrectly?
(serious operating error)
Mains voltage is 115 V, switch set
to 230 V?
Mains voltage is 230 V, switch set
to 115 V?
-
Change switch setting on the rear side
of the generator to 115 V.
Send in homogeniser for repair since
fault is serious.
Error analysis with HD 2070, HD 2200, HD 2070-U or HD 2200-U
Error Possible cause Remedial action
Device cannot be switched on?
Green control light does not light
up?
No power
Mains cable loose or faulty?
-
Check that plug connection is rmly in
place.
- Check the cable for continuity or, if
needed, exchange the mains connec-
tor.
Main fuse faulty?
- Replace main fuse.
No time display? Microcontroller defective? - Following unsuccessful display
test, see chapter 6.2 send
homogeniser for repair to the
manufacturer
30 / 44 51436a GB/2017-10