Instructions For Use
Guidance and Manufacturer’s Declaration – Emissions
The Site~Rite* 6 Ultrasound System is intended for use in the electromagnetic environment specified
below. The customer or user of the Site~Rite* 6 Ultrasound System should ensure that it is used in
such an environment.
Immunity Test EN/IEC 60601 Test
Level
Compliance
Level
Electromagnetic Enviroment –
Guidance
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3Vrms
3V/m
Portable and mobile communications
equipment should be separated from
the Site~Rite* 6 Ultrasound System by
no less than the distances calculated/
listed below:
D = 1.2 ( √ P)
D = 1.2 ( √ P) 80 to 800 MHz
D =
2.3(√P) 800 MHz to 2.5 GHz
wherePisthemaxpowerinwattsand
D is the recommended separation dis-
tance in meters.
Fieldstrengthsfromfixedtransmitters,
as determined by an electromagnetic
site survey, should be less than the
compliance levels.
Interference may occur in the vicinity of
equipment containing a transmitter.