38
Site~Rite
*
6 Ultrasound System
Recommended Separations Distances between portable and mobile RF Communications
equipment and the Site~Rite* 6 Ultrasound System
The Site~Rite* 6 Ultrasound System is intended for use in the electromagnetic environment in which radi-
ated disturbances are controlled. The customer or user of the Site-Rite* 6 Ultrasound System can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
Communications Equipment and the Site~Rite* 6 Ultrasound System as recommended below, according to
the maximum output power of the communications equipment.
MaxOutputPower
(Watts)
Separation (m)
150 kHz to 80 MHz
D = 1.2( √ P)
Separation (m)
80 kHz to 800 MHz
D = 1.2( √ P)
Separation (m)
800 MHz to 2.5 GHz
D=2.3(√P)
0.01 .1166 .1166 .2333
0.1 .3689 .3689 .7378
1 1.1666 1.1666 2.3333
10 3.6893 3.6893 7.3786
100 11.6666 11.6666 23.3333