Carbohydrate Labeling and Analysis Kit
Materials and Reagents
General Precautions
• Prior to using the instrument, turn on the laser and let it stabilize for 15 minutes. Turn it off
when not in use for longer periods of time.
• Do not interchange the capillary (i.e. if a capillary is being used for carbohydrate analysis, do
not use the same capillary in another application).
• When the capillary is not in use or after finishing with all experiments, it is recommended to
run a shutdown method to rinse the capillary and leave the ends immersed in water, to avoid
capillary blockage and to keep the capillary from drying out.
• Prior to using the capillary for the first time, it is recommended to rinse the capillary for 10
minutes at 30 psi pressure with DDI water and then rinse for 10 minutes at 30 psi with the
Carbohydrate Separation Gel Buffer.
Materials and Reagents
Contents of this Kit
Materials Needed but Not Provided in Kit
Table 1 Kit Contents Reorder Number (477600)
Component Quantity Reorder # Code
Carbohydrate Separation Buffer 56 mL 477623
N-CHO Coated Capillary 2 477601
Labeling Dye (APTS) 4x5 mg 501309 L6
Labeling Dye Solvent 1 mL L3
Glucose Ladder Standard 50 mg G20
Quantitation/Mobility Marker (Maltose) 0.18 mg G22
APTS-M (monosaccharide grade) 20 mg 725898
Description Part Number
LIF Performance Test Mix 726022
Universal Plastic Vials (pack 100) A62251
Universal Rubber Vial Caps - blue (pack 100) A62250
200
μL Micro Vials (pack 50)
144709
Assorted pipettors and pipette tips
Spatula
Analytical scale
Vortexer