Alarms
Precellys
®
24
Page 23/35
7 Alarms
WARNING:
Do not open the cover when the sample holder is moving.
7.1 Alarm Messages before Starting a Run
When the user presses the “Valid” button to start a run, the system checks information from the
different sensors before allowing a run to start:
Pressure sensor,
Cover sensor.
When the pressure is insufficient, the following message is displayed on the screen:
PRESSURE ALARM
CHECK SYSTEM
Red light is flashing / green light is off
If the cover is not closed properly, the following message is displayed on the screen:
TOP ALARM
CHECK SYSTEM
Red light is flashing / green light is off