Adjusting Parameters
Precellys
®
24
Page 22/35
6.3 Operating Ranges for Parameter Adjustments
In order to assure the optimal operation of this device, it is possible to use the parameters indicated
in the table below:
Speed (rpm)
Run Number Run Time
(s)
Pause (s) Waiting time between 2 cycles
5000 to 6200 1 to 3 5 to 90 5 to 120 2 to 5 mn
5 to 30 5 to 120
6300 to 6500 1 to 3
31 to 60 15 to 120
5mn
6600 to 6800 1 to 3 5 to 30 30 to 120 5 mn
Note:
A temperature sensor is used to automatically disconnect the engine power supply (instant stop
of the system) in case of important overheating. This alarm can be triggered during a run. When
the warning message appears on the display (see § 7.2.4), it is necessary to apply the procedure
described in § 7.3.