Maintenance
Precellys
®
24
Page 33/35
8.4 Cleaning and Decontamination
8.4.1 Recommendations
The casing of the unit can be cleaned up with a sponge or a damp cloth, moistened with water or
alcohol.
WARNINGS:
For safety purposes and to prevent any damage of the unit, the recommendations listed below
should be strictly followed:
Do not spray water or alcohol directly on the unit, especially in the air openings
and the fan.
Always disconnect the power cord before cleaning.
Do not use any type of scrapers.
Do not use caustic soda or acetone.
Do not use an aerial decontamination process.
8.4.2 Example of Decontamination Procedure
If a tube breaks when a run is in progress, decontaminate parts that may have been contaminated
with an appropriate disinfectant.
The decontamination procedure is of the sole responsibility of the user.
Parts that may have been contaminated can be cleaned with a sponge or a damp cloth moistened
with bleach at 6° Cl.
In case another decontamination procedure is to be applied, please first contact technical assistance
to ensure the compatibility of the new procedure with the instrument.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bertin Technologies PRECELLYS 24 and is the answer not in the manual?

Bertin Technologies PRECELLYS 24 Specifications

General IconGeneral
BrandBertin Technologies
ModelPRECELLYS 24
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals