Maintenance
Precellys
®
24
Page 29/35
8.2 Replacing Spare Parts
This paragraph lists the maintenance actions to be done by user on a regular basis, to ensure
PRECELLYS
®
24 runs properly.
Wearing parts are:
Wearing parts Reference
Frequency of
replacement
Why?
Indented plate 03119.810.NC002
1 year or to
destruction
Necessary to maintain tubes during
homogenisation
Toric joint 03119.810.NC003
6 months or to
destruction
Necessary to maintain tubes during
homogenisation
Containment seal 03119.810.NC004
1 year or to
destruction
Necessary to ensure the instrument’s
tightness
Fuse 03119.810.NC005 To destruction -
WARNING:
Potentially dangerous voltage exists inside the instrument. In order to ensure the user’s safety, the
containment seal should not be damaged in particular during the cleaning and/or decontamination
procedures.
Do not turn the unit upside down: the PRECELLYS
®
24 must always rest on its 4 feet for fear of
damaging internal components or breaking plastic casing.
8.2.1 Replacing Indented Plate
The indented plate can show signs of wear over time. Replace it once a year.