Maintenance
Precellys
®
24
Page 30/35
8.2.2 Replacing Toric Joint
The toric joint located on the holder can show signs of wear over time. It should be replaced when
the “Pressure Alarm” message is displayed permanently. It is however recommended to replace it at
least once every six months.
8.2.3 Replacing Containment Seal
The containment seal should be replaced at least once a year. It does not require the dismantling of
the holder.
WARNING:
Turn off the system and unplug power cord before replacing the containment seal.
1. Insert the interior part of the containment seal onto the metallic edge.
2. Insert the exterior part of the containment seal onto the casing.