3. Press "SETUP"
Screen reads; "Select FLOW RATE"
To change TARGET FLOW RATE;
* Press ENTER (Whole number value will remain on constant while the
tenths still blink); use "+" or "-" to increase or decrease until 16 is
displayed.
* Press ENTER (Tenths value will now remain constant while whole
number blinks); use "+" or "-" to increase or decrease until .7 is displayed.
Note: If the flow rate displayed is followed by Lpm then the selected flow rate is
calibrated and recalibration is not required. If CAL is displayed then calibration should
be performed as described in section 7.0.
4. Press "SETUP"
Screen reads; "Set DATE and TIME"
To change the Date and Time;
a. DAY: Press ENTER and change by pressing the + or - key.
When the day is correct, press ENTER.
b. MONTH: To change, press + or - key. When correct, press ENTER.
c. YEAR: To change, press + or - key. When correct, press ENTER.
d. TIME (hrs): To change, press + or - key. When correct, press ENTER.
e. TIME (min): To change, press + or - key. When correct, press ENTER.
5. Press "SETUP"
Screen reads; "Set START DATE and TIME"
This determines when you wish to start a sample job. It always defaults to
midnight the next day. Should you wish to change it;
a. DAY: Press ENTER and change by pressing the + or - key.
When the day is correct, press ENTER.
b. MONTH: To change, press + or - key. When correct, press ENTER.
c. YEAR: To change, press + or - key. When correct, press ENTER.
d. TIME (hrs): To change, press + or - key. When correct, press ENTER.
e. TIME (min): To change, press + or - key. When correct, press ENTER.
f. (enable): This is to turn this function On or Off. If "Off" then a sample job
(pump starts running) is initiated as soon as the "RUN/STOP" key
is pressed from the "MAIN IDLE DISPLAY". If "On" then the
sample job is initiated when the start date and time are achieved.
SET THIS FUNCTION TO "OFF" IF YOU WISH TO START
SAMPLING NOW. Change by pressing + or - key. When correct,
press ENTER. FOR THE PURPOSE OF THIS TUTORIAL, SET
TO "OFF".