6.7 Setting the Run Time
From the "MAIN IDLE DISPLAY" press the "Setup" key four times until the message
below appears;
Set RUN TIME
Hours:hh Min:mm eee
Where;
hh = hour (0-99 hours)
mm = minutes (0-59)
eee = enable "On" or "Off"
This sets the length of time that the sampler will run the pump for. If the setting is
acceptable, pressing the "Setup" key will change nothing and will step the menu to the
next selection. If a change is required, press the "Enter" key. You will notice that the
hours will then remain on while the other values continue flashing. The
NON-FLASHING value will now change when the "+/-" keys are pressed. When the
proper value has been achieved, pressing the "Enter" key will then select the next value
for editing. After all the values have been entered, pressing the "Setup" key will then
store the new values and step to the "MAIN IDLE DISPLAY".
This function can be disabled thereby allowing the pump to run until some other
shutdown such as Low Batteries etc. if the Enable (eee) is off.
IF YOU SIMPLY WANT THE PUMP TO RUN UNTIL SOME OTHER CONDITION
SUCH AS BATTERIES TOO LOW CAUSES A SHUTDOWN, THE LOWER RIGHT
OF THE DISPLAY SHOULD INDICATE OFF.
Note: When the reset key is pressed to clear the counters this setup will automatically
default to 24 Hours. If some other Run Time is required you must reset the run first
before using this Mode (See Resetting the Counters).
RESET DOES NOT AFFECT THE ENABLE AND WILL REMAIN UNCHANGED!
Note: If START TIME was enabled, after RUN TIME screen, the unit will not step the
screen to the AMAIN IDLE DISPLAY@. Instead, the next screen that appears will state;
To initiate a run,
Press Run/Stop