8
4.0 SPECIFICATIONS
4.1 Flow Rate Precision, Accuracy and Pulsation
Flow Rate Selections: 1.0 to 25.0 Lpm in .1 Lpm increments
Flow Rate Accuracy: .5%
(When calibrated with a DeltaCal or TetraCal (formally
tetraCal))
Flow Rate Precision: 2%
Pulsation for various applications as measured downstream from the filter:
Flow Rate Media Type Pulsation
------------ -------------- -----------
16.7 Lpm 47mm Quartz 4.56%
10.0 Lpm 47mm Quartz 3.84%
10.0 Lpm 47mm Teflo 7.38%
7.0 Lpm 47mm Quartz 5.06%
5.0 Lpm 47mm Quartz 6.13%
4.0 Lpm 47mm Quartz 7.94%
2.2 Lpm 37mm MCE w/BGI-4 Cyclone 4.70%
4.2 Run Times for Various Applications
Flow Rate Media Type Pressure Drop Run Time
------------ -------------- ---------------- ------------
25.0 Lpm 47mm Teflo 28.0cm H2O 11.6 Hours
25.0 Lpm 47mm Fiberglass 37.0cm H2O 7.4 Hours
16.7 Lpm 47mm Teflo 10.8cm H2O 33.6 Hours
16.7 Lpm 47mm Fiberglass 25.4cm H2O 32.4 Hours
16.7 Lpm 47mm Quartz 33.0cm H2O 28.4 Hours
14.0 Lpm 37mm MCE .8 112.0cm H2O 13.0 Hours
12.0 Lpm 37mm MCE .8 94.0cm H2O 13.5 Hours
12.0 Lpm 25mm MCE .8 170.0cm H2O 11.6 Hours
10.0 Lpm 37mm MCE .8 66.0cm H2O 19.8 Hours
2.2 Lpm 37mm w/BGI-4 Cyclone 5.0cm H20 67.0 Hours