43
06 Powered Off Normally! PQ Software
* Indicates that the PQ100 had last been powered off by pressing the "ON/OFF" button.
If you simply allow the PQ100 to power itself off (automatic after approximately 3
minutes of no user activity such as key presses or downloads) then the last Shutdown
code is retained.
07 Waiting for Alarm Event! PQ Software
* Indicates that the PQ100 has been programmed to start a sampling job at some future
time and is waiting for this occurrence.
08 MAXIMUM LOAD EXCEEDED! PQ100
Maximum Load Exceeded! PQ Software
* Indicates that the sample job was terminated because the PQ100 could not force the
pump to run any faster to regulate flow rate. This can happen when a large load pressure
is developed and the battery voltage drops. To eliminate this problem you can run from
the charger, recharge the batteries or try a different filter media.
09 Run Terminated by Operator! PQ Software
* Indicates that the sample job was terminated when an operator has pressed the
"RUN/STOP" button while a sample job was in progress.
10 RUN TIME EXCEEDED! PQ100
* Indicates that the PQ100 can not initiate a run as the RUN TIME (See Section 6.7) has
not been reset.
11 REPLACE CLOCK BATTERY! PQ100
* Indicates that the internal clock battery now requires replacement (See Appendix C).