Operating Manual
APT.line™ KMF (E5.2)
Constant climate chambers
with enlarged temperature and humidity range
with program control
Model
Art. No.
KMF 115 (E5.2)
9020-0187, 9120-0187
KMF 115-UL (E5.2)
9020-0188, 9120-0188
KMF 240 (E5.2)
9020-0145, 9120-0145
9020-0219, 9120-0219
KMF 240-UL (E5.2)
9020-0182, 9120-0182
KMF 720 (E5.2)
9020-0185, 9120-0185
9020-0245, 9120-0245
KMF 720-UL (E5.2)
9020-0186, 9120-0186
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 100
Internet http://www.binder-world.com
E-mail info@binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail service@binder-world.com
Service Hotline USA +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS +7 495 988 15 16

Table of Contents

Questions and Answers

  • K
    keith94Sep 8, 2025
    Why is there condensation at the walls of the inner chamber in my Binder KMF-240 Laboratory Equipment?
    • A
      Alex RoseSep 8, 2025
      If you observe condensation at the walls of the inner chamber of your Binder Laboratory Equipment, it could be due to a combination of temperature/humidity values not being in the optimum range; select a combination of temperature/humidity values in the optimum range. Alternatively, the set-point may have been too long below ambient temperature, causing icing in the preheating chamber; defrost the unit.
  • K
    Katie MedinaSep 7, 2025
    Why is the refrigerating performance low in my Binder KMF-240?
    • M
      Mark HicksSep 7, 2025
      If you are experiencing low or no refrigerating performance in your Binder Laboratory Equipment, it could be due to several reasons. The ambient temperature might be higher than 25 °C / 77 °F, in which case you should select a cooler place of installation. It could also be that the combination of temperature/humidity values is not in the optimum range; select a combination of temperature/humidity values in the optimum range. If the compressor is not turned on, contact BINDER service. Also, too much external heat load could be the cause, so reduce heat load.
  • M
    Michael LittleAug 22, 2025
    What to do if my Binder KMF-240 Laboratory Equipment chamber doesn’t heat up when turned on and the safety controller responds?
    • S
      Sabrina MarshallAug 22, 2025
      If your Binder Laboratory Equipment unit isn't heating up when turned on and the safety controller responds, it could be due to the limit temperature being reached. The safety controller might be set too low. To fix this, let the chamber cool down and hit the RESET button of the MB1 controller. If appropriate, select a suitable limit value. If the safety controller is defective, contact BINDER service.
  • D
    David McdonaldSep 4, 2025
    What to do if mechanical safety device class 3.2 responds in my Binder Laboratory Equipment?
    • M
      Mrs. Elizabeth PetersonSep 4, 2025
      If the mechanical safety device class 3.2 responds in your Binder Laboratory Equipment (with option safety device class 3.3), it could be because the limit temperature has been reached. Check the setting of the temperature set-point and of safety device class 3.2. If appropriate, select a suitable limit value. If the controller or safety device is defective, contact BINDER service.
  • M
    megan75Sep 3, 2025
    What to do if mechanical safety device class 3.1 responds in my Binder KMF-240?
    • E
      Elizabeth PattersonSep 3, 2025
      If the mechanical safety device class 3.1 responds in your Binder Laboratory Equipment (with option safety device class 3.3), it might be because the limit temperature has been reached. Check the setting of the temperature set-point and of safety device class 3.1. If appropriate, select a suitable limit value. Also, too much external heat load could be the cause, so reduce the heat load. If the controller is defective, contact BINDER service.
  • J
    jennifer65Sep 1, 2025
    Why my Binder KMF-240 Laboratory Equipment has no function?
    • J
      Joshua GonzalesSep 1, 2025
      If your Binder Laboratory Equipment unit has no function at all, start by checking the connection to the power supply. Also, verify that the power supply voltage matches the unit's requirement of 115V or 230V. Inspect the unit fuse and replace it if necessary. If the fuse responds again, contact BINDER service. A defective controller may also cause this issue, requiring you to contact BINDER service.
  • S
    swrightAug 31, 2025
    What to do if my Binder KMF-240 has no unit function and a dark display?
    • J
      jameshawkinsAug 31, 2025
      If your Binder Laboratory Equipment unit shows no function with a dark display, it could be due to the display mode being in “Standby”. To resolve this, simply press any controller key. Alternatively, the main power switch might be turned off; in that case, turn on the main power switch.
  • C
    Charles WilsonAug 30, 2025
    Why am I getting a humidity alarm message when operating without humidity on my Binder Laboratory Equipment?
    • R
      Rebecca MillerAug 30, 2025
      If you receive a humidity alarm message when operating without humidity (humidity switch (4) OFF) on your Binder Laboratory Equipment, it means that the humidity set-point is set to a value > 0% r.F. In Manual Mode: Enter a humidity set-point 0% r.H. In Program Mode: Enter a humidity subprogram with humidity set-point 0% r.H.
  • S
    Seth ColeAug 28, 2025
    How to fix low humidity and temperature accuracy in Binder Laboratory Equipment?
    • R
      raymondjacobAug 29, 2025
      If you're experiencing low humidity and temperature accuracy with your Binder Laboratory Equipment, it might be because the fan speed has been reduced. Set the fan speed to 100%.
  • B
    Brian BlackwellAug 27, 2025
    Why does my Binder KMF-240 have humidity fluctuation?
    • J
      Joshua FoxAug 27, 2025
      If you're experiencing humidity fluctuation and the control accuracy of ± 3 % r.F. is not reached in your Binder Laboratory Equipment, it could be due to a defective door gasket; replace the door gasket. Also, opening the door very frequently can cause this issue; try to open the doors less frequently.

Summary

EC - declaration of conformity

1. SAFETY

1.2 Structure of the safety instructions

Explanation of safety symbols and signal words used.

1.6 Intended use

Specifies suitable applications and restrictions for the chamber.

2. UNIT DESCRIPTION

4. INSTALLATION AND CONNECTIONS

4.4 Electrical connection

Instructions for connecting the unit to power.

5. START UP

5.1 Function overview of the MB1 display program controller

Explanation of the controller's interface and functions.

6. MB1 CONTROLLER SETTINGS

6.1 Selection of the MB1 program controllers menu language

How to change the controller's display language.

6.2 Function overview of the MB1 program controller displays

Description of the different display modes.

8. MANUAL MODE

8.1 Entering the set point values

How to input temperature and humidity set points.

9. PROGRAM OPERATION

9.1 Menu-based program entry

Steps to create programs using the controller menu.

9.2 Entry of temperature values and fan speed

How to input temperature and fan speed in programs.

9.3 Entry of humidity values

How to input humidity values in programs.

10. TEMPERATURE SAFETY DEVICES

10.1 Over temperature protective device (class 1)

Description of the primary over-temperature protection.

10.2 Safety controller (temperature safety device class 3.1)

Description of the secondary safety temperature controller.

10.2.1 Safety controller set-point types

Explanation of 'Limit' and 'Offset' set-point types.

10.3 Temperature safety device class 3.3 (DIN 12880) (option)

Description of the optional over/under temperature protection.

10.3.1 Temperature safety device class 3.1

Detailed function and setting of class 3.1 safety device.

10.3.2 Temperature safety device class 3.2

Detailed function and setting of class 3.2 safety device.

11. NOTIFICATION AND ALARM FUNCTIONS

11.2 Resetting the notification or alarm messages

Procedure to acknowledge and clear alarms.

12. HUMIDITY SYSTEM

12.1 Function of the humidifying and dehumidifying system

Explanation of how the humidity system operates.

13. DEFROSTING AT REFRIGERATING OPERATION

15. MAINTENANCE, CLEANING, AND SERVICE

15.1 Maintenance intervals, service

Schedules and procedures for maintenance.

15.2 Cleaning and decontamination

Procedures for cleaning and decontaminating the unit.

17. TROUBLESHOOTING

Binder KMF-240 Specifications

General IconGeneral
Interior Volume240 liters
Temperature Variation±0.5°C
Max. Load per Shelf30 kg
Permitted Total Load80 kg
Nominal Voltage230 V
Power Supply230 V 50/60 Hz
Interior Dimensions (W x H x D)600 x 1400 x 600 mm
Interior Dimensions (W x D x H)600 x 600 x 1400 mm
Exterior Dimensions (W x D x H)800 x 800 x 1900 mm