KMF (E5.2) 06/2014 page 88/99
Unit size
(Art. No.)
115
(9x20-0187)
240
(9x20-0145)
240
(9x20-0219)
720
(9x20-0185)
(9x20-0245)
Electrical data
IP system of protection acc. to EN
60529
20 20 20 20
Nominal voltage (+/-10%)
V
200 to 240 1N~
Power frequency Hz 50/60 50/60 50/60
50/60
(9x20-0185)
50
(9x20-0245)
Power plug
shock proof plug
Nominal power
kW
2.00
2.10
2.50
3.10
Energy consumption 4) at 85 °C / 185
°F and 85 % r.H.
Wh/h 570 500 720 1050
Installation category acc. to IEC 61010-
1
II II II II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Over-current release category B, 2
poles
Amp 16 16 16 16
Noise level approx.
dB (A)
52
52
52
53
Electrical connection data KMF constructed acc. to UL requirements (for the USA and Canada)
Unit size
(Art. No.)
115
(9x20-0188)
240
(9x20-0182)
720
(9x20-0186)
Electrical data KMF-UL
IP system of protection acc. to EN 60529
20
20
20
Nominal voltage (+/-10%) 50/60 Hz
V
200 to 240 / 2~
Nominal power
kW
2.00
2.10
3.10
Power plug
NEMA
6-20P
6-20P
6-20P
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Over-current release category B, 2 poles
Amp
16
16
16
1) Lower values are valid up to an ambient temperature of max. 25 °C / 77 °F
2) to 98 % of the set value
3) Depending on the set-point.
4) Use this value for sizing air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of +/-10%. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880, observing the recommended wall clear-
ances of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan
speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
Refrigerating performance decreases while operating the chamber at temperatures < 0 °C / 32
°F due to icing of the evaporators. For this reason defrost the chamber regularly, e.g. once a
week.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.