Manual for VMP300-based instruments
100
Impedance specifications (if booster board connected to board with EIS option)
Accuracy
10.2.6 10 A/ 5V booster
Control amplifier
Compliance voltage
-1V; +6V
Compliance current
± 10 A
Control voltage
-1V, 6V
Bandwidth (-3dB)
>8MHz
Operating area
Slew rate (no load)
>50V/µs
Rise/Fall time (no load)
<200ns
Floating mode
yes
Parallel ability
Yes
Current accuracy
0.1% range
Impedance specifications (if booster board connected to board with EIS option)
-4
-2
0
2
4
6
8
-12 -8 -4 0 4 8 12
E(V)
I(A)