LifePro Thermal Cycler Failure analysis and handling
-40-
Hot-lid closes
automatically since
temperature control
time is set as “--:--“ in
file editing interface
Set temperature
control time in digital
form
Hot plug-in is loose
Heating element in
hot-lid is damaged
Temperature sensor in
hot-lid is damaged
Contact supplier or
manufacturer
Chip is in poor contact
10 Unusual character is shown on screen
Chip is damaged
Contact supplier or
manufacturer
11 Button on panel can’t work
Micro-switch is
damaged
Contact supplier or
manufacturer
Hot-lid temperature is
not set. Hot-lid is set as
“OFF”
Set hot-lid
temperature as per
How to Set System
Parameters in 1.2 of
Chapter IV in this
manual
Reaction tube is not
placed uniformly
Adjust hole position
on reaction tube, try to
place symmetrically
12 Reagent in reaction tube evaporates
Reaction tube is not
covered tightly
Please cover reaction
tube tightly before put
into instrument
13 System parameter setting is invalid
It didn’t press “Save” to
confirm system
parameter setting
Please refer to 1.2 of
Chapter IV in this
manual
Caution: 1. It is forbidden for user to open the case and check by oneself in warranty period.
Please contact supplier or manufacturer in time if any failure needs to open the
case to check.
2. Since temperature control characteristics (rate of temperature rise/fall, stability,
fluctuation and etc.) of different brands and each instrument are specific, and
biotic experiment itself has uncertainty and vulnerable to outside influence, PCR
program that runs well on one instrument might fail to reach same effect on
another. Therefore, please adjust PCR running program to get ideal performance
once instrument is changed.