Patient Monitor Users Manual
5-2
When conducting defibrillation, it is imperative to
only use the ECG electrodes and cables specified by
manufacturer.
Check the ECG cable and its package for any sign of
damage before use. Do not use the cable if any damage
is detected.
Do not come into contact with the patient during
defibrillation. Otherwise serious injury or death could
result.
When the electrode polarized voltage is too high, the
monitor will indicate the abnormal state by alarm
system.
The monitor is protected against defibrillation effect.
When applying defibrillator to the patient, the
monitor will experience transient disorderly
waveforms. If the electrodes are used and placed
correctly, the display of the monitor will be restored
within 10s. During defibrillation, the chest leads such
as V1~V6 should be removed and such limb electrodes
as RA, LA, RL, LL should be moved to the side of the
limbs.
When you are connecting the electrodes or the patient