Patient Monitor Users Manual
7-4
7.3 Monitoring Procedure
1. Selecting SpO
2
Sensor
Depending on the patient category, weight and application
site, you can select the SpO
2
sensor as required.
2. Connecting SpO2 Sensor
Plug the SpO
2
sensor cable into the SpO
2
connector on the
measurement module.
3. Applying SpO2 Sensor
Clean the application site, such as colored nail polish, and
apply the sensor to the patient
Warning:
Do not use the SpO2 sensor on a limb where the NIBP
cuff is applied. This may result in inaccurate SpO2
reading due to blocked blood flow during cuff
inflation.
Do not conduct SpO2 measurement on the finger
smeared with nail polish, otherwise unreliable
measurement results might be produced.
When using finger sensor, make sure the nail faces to
the light window
0