7-8
7.6.1 SpO2AlarmSetup
Click the right of Alarm, you can set alarm level of SpO
2
and PR, you can select “Mid, High
7.6.2 SpO2AlarmLimitsSetup
Click the right of Uplimitor Downlimit, you can set
the SpO2 uplimit and downlimit. Attention: The high alarm limit
should greater than the low one.
7.6.3 ScanSpeedSetup
Click the right of Speed, you can select “6.25 mm/s
12.5mm/s, 25 mm/s”.
7.6.4 AverageTimeSetup
Click the right ofAvg Time, you can select “4s, 8s, 16s”.
7.6.5 QRSVolumeSetup
Click the right ofQRS Vol., you can selectOff, High,
Mid, Low”.
7.7 Desat limit Setup
SpO
2
desat means when SpO
2
measuring value is lower than