Patient Monitor User’s Manual
5
Patient ID 1~96
Display way Trend tabular
trend interval 10s, 30s, 1min, 2min, 5min, 10min
Storage Save when power down
Capacity
500 groups/patient can be stored (only
data, no waveform).
A.7 Measurement Specifications
A.7.1 ECG Specifications
ECG
Lead type 3-lead:I,II,III
5-lead:I, II, III, aVR, aVL, aVF, Vx
Lead standard AHA、IEC
ECG input
rang
±6mV
Gain 2.5mm/mV, 5mm/mV, 10mm/mV
Input
impedance
≥5.0MΩ
Leakage
Current
<10 uA