Patient Monitor Users Manual
5-3
cable, make sure that the connectors never come into
contact with other conductive parts, or with earth. In
particular, make sure that all of the ECG electrodes
are attached to the patient, to prevent them from
contacting conductive parts or earth.
Please check the skin where the electrodes are placed,
replace the electrodes or relocate the electrodes in case
of skin allergy occurs.
Interference from instruments near the patient and
ESU interference can cause problems with the ECG
wave.
The monitor cannot be directly applied to heart and
cannot be used for the measurement of endocardio
ECG.
Line isolation monitor transients may resemble actual
cardiac waveforms and thus inhibit heart rate alarm,
at this time, you should unplug power adapter and
supply power by lithium battery.
5.3 Monitoring Procedure
5.3.1 SkinPreparationforElectrodePlacement