EasyManua.ls Logo

Biolight Meditech M850 - Page 78

Default Icon
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor Users Manual
10-3
functional damage.
Inspect the safety relevant labels for legibility.
Verify that the device functions properly as described in the
instructions for use.
Test the earth leakage current according IEC 60601-1 Limit:
NC 500μA, SFC: 1000μA.
Test the enclosure leakage current according to IEC 60601-1:
Limit: NC 100μA, SFC: 500μA.
Test the patient leakage current (normal operation) according
IEC 60601-1
Limit: type CF: for a.c.: 10μA, for d.c.: 10μA.
Test the patient leakage current under single fault condition
according IEC 60601-1
Limit: type CF: for a.c.: 50μA, for d.c.: 50μA.
Test the patient leakage current Mains voltage on applied
part: According IEC 60601-1:
Limit: type CF: for a.c.: 50uA.
Warning: No use-serviceable parts inside, before
servicing to authorized representative or
manufacturer.
0

Table of Contents

Related product manuals