74
Calibration and Wi-Fi Connection(Optional)
2. Enter the exact BG value then press [Next]. If incorrect, tap the backspace key on the numeric
keypad to erase the entries then enter the correct value.
3. You will see a prompt from receiver. Tap [Yes] or [No].
Wi-Fi CONNECTION (for HR321)
Receiver could access to internet via Wi-Fi network. Follow the instuction to set the Wi-Fi connection
in Setting.
NOTE :
Make sure you enter exact fingerstick blood glucose value that is measured within 5 minutes.
Only a BG value between 40 mg/dL (2.2 mmol/L) and 500 mg/dL (27.8 mmol/L) can be
accepted for calibration. If your BG value is significantly different from your sensor reading, it
is recommended to calibrate again to avoid inaccurate readings.
2
75
Calibration and Wi-Fi Connection(Optional)
3