EasyManua.ls Logo

BIONIME RIGHTEST - XII. Rightest Cgms Specifications; Sensor Inserter Specifications

Default Icon
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84
Specifications
XII. RIGHTEST CGMS SPECIFICATIONS
SENSOR INSERTER SPECIFICATIONS
Sensor Glucose Range
Operating and Storage Altitude
40 - 500 mg/dL
0 to 3,048 metres (0 to 10,000 ft)
Sensor Useful Life up to 14 days
Sensor Inserter Shelf Life 12 months
Sensor Operating Conditions
Temperature: 5°C - 45°C (41°F - 113°F)
Relative humidity: 10% - 90%
Sensor Ingress Protection Rating
(when installed with a transmitter)
Inserter Storage & Transportation
Conditions
IP47
Protected from tools and small wires greater than 1 millimeter.
Protected from immersion between 15 centimeters and 1
meter in depth.
Temperature: 5°C - 30°C (41°F - 86°F)
Relative humidity: 10% - 90% (in a cool, dry place)
Inserter Size 52.0 x 57.0 x 61.3 mm (± 0.5 mm)
Sterilization
Usage
Sterilized by radiation
Single use (disposable)
85