Appendix
XIV. APPENDIX: ALARM/ALERTS
RECEIVER : GLUCOSE AND SIGNAL LOSS ALERTS
Screen Event Visual & Sound Settings
Very High Alert
happens when glucose
> 250 mg/dL and is detected to
rise or remain steady.
Default Setting:
Visual Safety Symbol ( ) on
Home Screen: Yes.
Sound & Vibration Alert: Yes.
- Vibrates 3 times and beeps 3
times at the alarm volume
following the user settings.
Separated by 2s.
- Sound & Vibration active
unless turned off by user
pressing the [OK] button on
screen; or glucose is projected
to fall.
- Very high glucose alert will
repeat every 30 minutes or
until glucose level returns to
the target range.
Editable Setting:
Sound & Vibration can be
turned ON/OFF by the user.
97
Appendix
Screen Event Visual & Sound Settings
Low Glucose Alert
happens when glucose is
between 55 - 70 mg/dL and is
detected to fall or remain steady
Default Setting:
Visual Safety Symbol ( ) on
Home Screen: Yes.
Sound & Vibration Alarm : Yes.
- Vibrates 2 times and beeps 2
times at the alarm volume
following the user settings.
Separated by 1s.
- Sound & Vibration active
unless turned off by user
pressing the [OK] button; or
glucose is projected to rise.
- Low glucose alert will repeat
every 30 minutes or until
glucose level returns to the
target range.
Editable Setting:
Sound & Vibration can be
turned ON/OFF by the user.