98
Appendix
Screen Event Visual & Sound Settings
Urgent Low Alarm
happens when glucose is
at or below 54 mg/dL.
Default Setting:
Visual Safety Symbol ( ) on
Home Screen: Yes.
Sound & Vibration Alarm : Yes.
- Vibrates 2 times and beeps 2
times at highest alarm volume.
Separated by 1/2s.
- Sound & Vibration active
unless turned off by user
pressing the [OK] button; or
until glucose is higher than
55 mg/dL.
- Urgent low glucose alarm will
repeat every 30 minutes or
until glucose level returns to
the target range.
Editable Setting:
Sound & Vibration CANNOT
be turned ON/OFF by the
user.
99
Appendix
Screen Event Visual & Sound Settings
Transmitter is too far from the
receiver or when there is an
obstacle (e.g. water, wall) in
between transmitter and
receiver .
Default Setting:
Visual message on the screen:
Yes
Sound & Vibration Alert : Yes.
- Vibrates 1 time and beeps 1
time at highest alarm volume.
Separated by 1s.
- Displayed continuously for 5
minutes or until [OK] is
pressed.
System attempts to reconnect
every 5 minutes even if user
doesn’t press [OK]
System alerts will repeat every
30 minutes.
Editable Setting:
Sound & Vibration CANNOT
be turned ON/OFF by the user.