114 115
Test
frequency
(MHz)
Band a)
(MHz)
Service a) Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
Compliance
LEVEL
(V/m)
(for home
healthcare)
710 704–787 LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2 0,3 9 9
745
780
810 1 700 –
1 990
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2 0,3 28 28
870
930
1 720 2 400 –
2 570
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2 0,3 28 28
1 845
1 970
Test
frequency
(MHz)
Band a)
(MHz)
Service a) Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
Compliance
LEVEL
(V/m)
(for home
healthcare)
2 450 2 400 –
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2 0,3 28 28
5 240 5 100 –
5 800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2 0,3 9 9
5 500
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does
not represent actual modulation, it would be worst case.