Page 20
nGEN™ Pump User Manual
12 Technical Data
12.1 Specifications
AC Adapter
AC input 100 to 240 VAC
Frequency 50 Hz / 60 Hz
DC output 24 V, 3 A
Flow Accuracy 1 to 5 ml/min -10% to +20%
6 to 39 ml/min -5% to +15%
40 to 60 ml/min -10% to +20%
Bubble Detection Designed to detect 0.4 µl or larger bubbles at the flow
rates specified below (equivalent to minimum sensitivity
of 2 µl bubbles at atmospheric pressure)
Adjustment of the Flow Rates
(in 1 ml/min steps)
Low flow 1 ml/min to 5 ml/min
High flow
4 ml/min to 60 ml/min
Flush Flow Rate 100 ml/min
Maximum Infusion Pressure
˂ 150
psi
Sound Pressure Level for Alarm Tone 55 to 65 dBA (measured 1 m from pump, factory
setting that cannot be changed by the user)
Classification Class I and defibrillation proof type CF (as defined in
IEC/EN 60601-1)
Class IIa
Weight 5.5 kg (14.7 lb)
Dimensions 19 cm deep x 18 cm wide x 28 cm high
(7.5 in x 7.1 in x 11.0 in)
IP Rating IP23
AVISO: Este es un documento controlado, confidencial, y con derechos reservados. Revisar si es la revision mas actualizada.
WARNING: This is a controlled proprietary and confidential document. Verify revision is current prior to use.
Release Date: 2/14/2024