nGEN™ Pump User Manual
12.2 Operating, Storage, and Shipping Specifications
Minimum Maximum
Operating Specifications
Ambient temperature 10°C 40°C
Relative humidity 10% 90%, non-condensing
Atmospheric pressure 700 mbar 1060 mbar
Storage Specifications
Ambient temperature -30°C 65°C
Relative humidity 0% 90%, non-condensing
Atmospheric pressure 500 mbar 1060 mbar
Shipping Specifications
Ambient temperature -30°C 65°C
Relative humidity 0% 90%, non-condensing
Atmospheric pressure 500 mbar 1060 mbar
12.3 EMC Information and Technical Description
The use of accessories and cables other than those specified in this user manual or specified by
Biosense Webster may result in increased emissions or decreased immunity of the pump.
The pump is intended for use in the electromagnetic environment specified in Table 7 through Table
10. The user of the pump should ensure that it is used in such an environment.
A false bubble alarm might occur if the pump is operated in an environment where the
electromagnetic disturbances exceed those specified in the tables below. For information on alarms,
see 10.10, Alarms.
Table 7 – Electromagnetic Emissions
Emissions Test Compliance Electromagnetic Environment –
Guidance
RF emissions CISPR 11 Group 1 The pump uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11 Class A
The pump is suitable for use in
establishments other than domestic and
those directly connected to the public
low-voltage power supply network.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
AVISO: Este es un documento controlado, confidencial, y con derechos reservados. Revisar si es la revision mas actualizada.
WARNING: This is a controlled proprietary and confidential document. Verify revision is current prior to use.
Release Date: 2/14/2024