Page 23
nGEN™ Pump User Manual
Table 9 Electromagnetic Immunity
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
3 V
0.15MHz to
80 MHz
6 V in ISM
bands between
0.15 MHz and
80 MHz
80% AM at
1 kHz
3 V
0.15 MHz to
80 MHz
6 V in ISM
bands between
0.15 MHz and
80 MHz
80% AM at
1 kHz
Portable and mobile RF communications
equipment should be used no closer to any part of
the pump, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.7 GHz
80% AM at
1 kHz
Test frequency
385 MHz to
5.750 GHz with
pulse
modulation
3 V/m
80 MHz to
2.7 GHz
80% AM at
1 kHz
Test frequency
385 MHz to
5.750 GHz with
pulse
modulation
Recommended separation distance:
Pd 17.1
150 kHz to 80 MHz
Pd 17.1
80 MHz to 800 MHz
Pd 23.2
800 MHz to 2.7 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
pump is used exceeds the applicable RF compliance level above, the pump should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.