EasyManua.ls Logo

BIRD 4391A - Initial Setup

BIRD 4391A
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
18
Initial Setup
Note: On a clean flat workspace perform the following setup pro-
cedure:
1. Remove the front panel. Refer to
"Front Panel Removal" on page 21
.
2. Insert a DC feed-in adapter into the FWD element of the socket.
3. Insert a metal dust plug (supplied with the instrument) into the RFL
element socket.
4. Connect a BNC to BNC cable assembly from the output of calibrator/source
to the DC feed-in adapter.
5. Set the RF Power Analyst range switches to the 10 and x100 positions.
6. Connect AC power and place the LINE/OFF/BATT switch to the LINE setting.
7. Switch on the Calibrator/Source.
8. Allow it to warm up for at least one hour.
9. Set the Calibrator/Source to output 30 ±0.05 microamps.
10. Adjust potentiometer, R25 (
Figure 15 on page 18
), until the RF Power
Analyst Displays 1000 ±5.
Figure 15 Calibration Potentiometer

Related product manuals