EasyManua.ls Logo

BK Precision 4052 - Troubleshooting, Specifications, Service, and Warranty; Troubleshooting Common Issues

BK Precision 4052
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70
4 Troubleshooting Guide
Below are some frequently asked questions and answers. Please check if any apply to your
power supply before contacting B&K Precision.
Q: I cannot power up the generator
- Check that the power cord is securely connected to the AC input and there is live power
from your electrical AC outlet.
- Verify that the AC power coming from the mains is the correct voltage. The generator
can accept a specific range of AC input voltages. Refer to section “2.1”.
Q: I cannot see any output out of the output terminals
- Make sure that the Output buttons above channel 1 and/or channel 2 is lit. If not,
press it once. The backlight of these Output buttons indicates that the respective
channel’s output is enabled.
Q: I have connected my signal to an oscilloscope, but the amplitude is double of what I set
- This is often because the impedance of the generator is not matched with the
oscilloscope. When the generator is set to 50 impedance, connecting it directly to an
oscilloscope with a 1 M input impedance will cause this to happen. To eliminate this
issue, connect a 50 terminator to the input of the oscilloscope, then connect a BNC
cable between the generator and the terminator to have matching impedance.

Related product manuals