EasyManua.ls Logo

Bodystat 1500 - EMC - ELECTROMAGNETIC COMPATIBILTY

Bodystat 1500
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
EMC ELECTROMAGNETIC COMPATIBILTY
Table 1 - Guidance and manufacturers declaration electromagnetic
immunity
The Bodystat 1500 is intended for use in the electromagnetic environment
specified below. The customer or the user of the Bodystat 1500 should assure
that it is used in such an environment.
Emissions test
Compliance
RF Emissions
CISPR 11
EN 55011: 2016 +
A11: 2020
Group 1
The 1500 must emit electromagnetic
energy in order to perform its intended
function. Nearby electronic equipment
may be affected
RF Emissions
CISPR 11
EN 55011: 2016 +
A11: 2020
Class A
The Bodystat 1500 is suitable for use
in all establishments other than
domestic and those directly connected
to the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
EN 61000-3-2: 2019
Not
applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
EN 61000-3-3: 2013
+A1:2019
Not
applicable