Table 3 - Guidance and manufacturers declaration electromagnetic immunity
The Bodystat 1500 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Bodystat 1500 should assure that it is used in such an environment
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
EN 61000-4-3:
2020
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,7 GHz
27 V/m, 18 Hz PM
385 MHz
28 V/m, 50 %18 Hz PM
450 MHz
9 V/m, 217 Hz PM
710 MHz
9 V/m, 217 Hz PM
745 MHz
9 V/m, 217 Hz PM
780 MHz
28V/m, 18 Hz PM
810 MHz
28 V/m, 18 Hz PM
870 MHz
28 V/m, 18 Hz PM
930 MHz
28V/m, 217 Hz PM
1720 MHz
28 V/m, 217 Hz PM
1845 MHz
28 V/m, 217 Hz PM
1970 MHz
27 V/m, 217 Hz PM
Not applicable
3 V/m
80 MHz to 2,7 GHz
27 V/m, 18 Hz PM
385 MHz
28 V/m, 50 %18 Hz PM
450 MHz
9 V/m, 217 Hz PM
710 MHz
9 V/m, 217 Hz PM
745 MHz
9 V/m, 217 Hz PM
780 MHz
28V/m, 18 Hz PM
810 MHz
28 V/m, 18 Hz PM
870 MHz
28 V/m, 18 Hz PM
930 MHz
28V/m, 217 Hz PM
1720 MHz
28 V/m, 217 Hz PM
1845 MHz
28 V/m, 217 Hz PM
1970 MHz
27 V/m, 217 Hz PM
Were P is maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended
separation distance in metres
(m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol: