EasyManua.ls Logo

Braun ThermoScan PRO 6000 - Cleaning and Disinfecting the Thermometer Body and Cradle

Braun ThermoScan PRO 6000
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English
4
With probe facing down, wipe the probe with a damp cloth or
cleaning wipe moistened with isopropyl or ethyl alcohol.
5
Gently wipe with dry, clean cotton swab or cloth immediately.
6
Allow at least 5 minutes drying time before taking temperature.
Make sure probe lens window is clean and dry prior to use.
14.2 Cleaning and disinfecting the thermometer body and cradle
Do not submerge the thermometer. Excess liquid may cause damage to the thermometer.
Wipes should be moist, not saturated.
Do not use any chemical other than those listed in the Approved Cleaning Solution Table to clean
the thermometer body and cradle. Other cleaning agents could cause damage to the thermometer.
When cleaning the probe lens window or probe ONLY use isopropyl or ethyl alcohol.
Do not use abrasives pads or cleaners.
Approved cleaning solutions
Family Solution or
Brand
Probe Lens
Window
Probe Thermometer
Body & Cradle
Tether
Chlorine
and Chlorine
Compounds
10% Chlorine
Bleach solution
No No Yes Yes
Quaternary
Ammonium
Compounds
CaviWipes
Clinell
®
Universal Wipes
SaniCloth
Metrex
No No Yes Yes
Hydrogen
Peroxide
Virox
Oxivir
No No Yes Yes
Alcohol 70% isopropyl
or ethyl alcohol
Yes Yes Yes Yes
Additional cleaning agents may be periodically evaluated for compatibility. If your cleaning agent is
not listed, contact Welch Allyn to determine if additional cleaning agents are approved for use.